high operating pressures to avoid filter burst AP-4307 2.1 bar at 21-24 °C (70-75 °F) at 2.1 bar ... Acrodisc PSF GxF/GHP V o l u m e F i l t e r e d (m L)
3.9 cm² Filter Media Glass Fiber / GH Polypro membrane (GHP, hydrophilic polypropylene) Hold-Up Volume <200 µL Housing Material Polypropylene Maximum Operating Pressure (bar) 4.1 bar Maximum Operating Temperature (Metric) 55 °C Typical Flow Rate 175 mL/min at 2.1 bar (210 kPa, 30 psi) Overview Description Type Use
25 mm Effective Filtration Area (Metric) 2.8 cm² Filter Media Nylon Hold-Up Volume <125 µL Housing Material Polypropylene Maximum Operating Pressure (bar) 2.1 bar Maximum Operating Temperature (Metric) 55 °C Typical Flow Rate 245 mL/min at 2.1 bar (210 kPa, 30 psi) Overview Description Applications Type Use Related Products
Pall Acrodisc Premium Syringe Filters, Pall Life Sciences AP-4438, Case of 200 ON SALE . Pall Laboratory Filters, Filters, Syringe filters. +1-800-504-7309 Contact Check Order Status
no leakage was observed the pressure was slowly ramped up to 3.4 barg ( 50 psig). No leakage or water breakthrough was observed up to a test pressure of 3.4 barg (50 psig). The test results demonstrate that the capsules retain their hydrophobic property and do not allow water to break through up to a pressure of 2.1 bar (30 psi).
Acrodisc® Syringe Filters, 13mm, Pall Gelman Taylor Scientific 12-4616 12-4616-01 12-4616-02 12-4616-03 12-4616-04 Acrodisc Syringe Filters, 13mm, Pall Gelman 4540
Acrodisc Syringe Filters with GHP Membrane 55 C (131 F) at 2.1 bar (210 kPa, 30 psi) Maximum Operating Pressure 13 mm: 6.3 bar ... Popular in High Performance ...
Acrodisc® Syringe Filters with GHP Membrane - Acrodisc® Syringe Filters at 2.1 bar (210 kPa, 30 psi) Maximum Operating Pressure. 13 mm: 6.3 bar (630 kPa, 90 psi ...
30 Acrodisc PSF (25 mm Syringe Filter) Product Specifications Typical Hold-Up Maximum Maximum Typical Part Pore Volume Operating Operating Water Number Description Size (w/air purge) Temperature Pressure Flow Rate AP-4305 GxF/GHP 0.2 µm < 200 µL 55 °C (131 °F) at 2.1 bar 4.1 bar (410 kPa, 60 psi) 175 mL/min at 2.1 bar AP-4307 (210 kPa, 30
Maximum Operating Temperature 55 °C (131 °F) Maximum Operating Pressure 6.2 bar (620 kPa, 90 psi) Typical Water Flow Rate 60 mL/min at 2.1 bar (210 kPa, 30 psi) Endotoxin Level < 0.25 EU/mL using Limulus Amoebocyte Lysate (LAL) test . Sterilization Sterilized by ethylene oxide and individually packaged; Additional Information: Item Code: 4433
Acrodisc Syringe Filters with GHP Membrane. Filter Media: Maximum Operating Pressure: ... Typical Water Flow Rate (mL/min at 2.1 bar (30psi)) 0.2µm, 25mm is 60 ...
BSA: 9 – 11 mg/Acrodisc or 50 – 60 mg/mL membrane volume Maximum Operating Pressure and Temperature: 5.5 bar (80 psi) at 21 – 24 ºC 2.1 bar (30 psi) at 60 ...
Burst/Pressure Test Acrodisc syringe filters are tested for resistance to pressure. This test is a quality and safety test to ensure the Acrodisc syringe filter housings will not rupture at their rated operating pressures. The maximum operating pressure listed in the product literature is well below the actual burst pressure.
Maximum Operating Pressure (bar) 2.1 bar Maximum Operating Temperature (Metric) 55°C Typical Flow Rate 215 mL/min at 2.1 bar (210 kPa, 30 psi) Overview Description Type Use Documents Datasheets Download: English Description Syringe filters are used in analytical chemistry applications to remove particulate contamination prior to analysis.
PALL AP-4560 Acrodisc PSF Syringe Filter with GHP Best Sellers Customer Service Prime New Releases Today's ... 55°C Maximum Operating Temperature At 2.1 Bar